LDT syringe : refusal of the ISO 20695 committee to include the Low Dose Tip (LDT) syringe in the normative requirements
The ISO committee decided to remove the Low Dose Tip (LDT) syringe from the normative requirements of ISO 20695 (EN1615/EN1618 revision, specifying the enteral feeding devices requirements). The main reason is that the Low Dose Tip” (LDT) syringe can’t ensure to reliably increase dose accuracy.